Instruction Manual ColorPlus 2
Commissioning
52
13045E/3
6
Commissioning
The initial start-up of the web user interface via the Ethernet interface is described in the
Reference Handbook.
Consult the corresponding documentation when using a SICON M (document no.
11775 / 11776).
If malfunctions occur, consult the Section 10.
Proceed with the initial start-up in accordance with the following table:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Ensure that the photometer and control unit
are correctly mounted and connected.
2.
2.1: Establish the service voltage to the pho-
tometer and switch it on. The welcome screen
appears on the display.
The factory setting language is English.
Accordingly, the displayed language during the
initial start-up is English.
2.2: The instrument carries out an internal
functional check.
2.3: The instrument is ready for measurement.
3.
Set the language.
Section 8.1
4.
Set the current outputs.
Section 8.2
5.
Set the limits.
Section 8.3
6.
Set the outputs.
Section 8.4
7.
Set the optional functions according to the
Reference Handbook.
For example, cleaning cycle with
ColorPlus 2, analog input compen-
sation