Instruction Manual ColorPlus 2
Mounting
32
13045E/3
4.2.2
Mounting the photometer on the inline housing
Figure 8 : Cross-section of ColorPlus 2
Transmitter
Ring adapter
Measuring cell window
Inline housing
Receiver with external checking
unit
OPL bit
Lock ring
Flood protection
The ColorPlus 2 is delivered with an OPL bit (Figure 8, pos. 6) on both sides for installation in
an inline housing. The OPL bits with measuring cell window (Figure 8, pos. 3), the corre-
sponding flood protection (Figure 8, pos. 8) and ring adapter (Figure 8, pos. 2) are already
installed according to customer requirements at the factory.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Insert the complete transmitter/receiver into
the mounting on the inline housing.
The groove (X) on the transmitter and
receiver must face forwards when mounted.
2.
Loosely position the transmitter/receiver on the
inline housing with the lock ring (Figure 8, pos.
7) and tighten with the nut (A).