
Operation
5.11 Configuration
FIDAMAT 6-G
5-62
Operating instructions, 07/2006, A5E00478463
5.11.20
Start-up state (function 91)
91 Startup status
THC
Pause
Pause without heating
Figure 5-69 Changing the start-up state / operating state
With this function, you put the device into different start-up states (operating state to which
the device switches after it has successfully run through the warm-up phase). Also, you can
exit certain operating states with this function, for example those which are brought about by
faults/maintenance requests. The respectively currently active operating state is marked with
"■", and the others with "□".
Pause
With this setting, you reduce the gas consumption.
Pause without heating
This setting has the same configuration as the "pause" mode, but the heating is switched off
as well.
Note
Changing the operating state
The operating state can either be changed manually using function 91, or automatically by
occurring faults / maintenance requests. The table “Change of state following
error/maintenance request” shows possible state changes.
See also