Operation
5.11 Configuration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-41
Function
Relay is passive when Relay conducts current Remark
Sample gas
Feed of sample gas
Valve actuation
Zero gas 1
Zero gas feeding
Test gas 1 to 4
Feed of test gas
Measuring point 1 to 6
Measuring point 1 to 6
activated
For taking a gas
sample via solenoid
valves at various
measuring points
Measuring point signal
1 to 6
Measuring point 1 to 6
activated
For identifying the
measuring point (runs
parallel to measuring
point)
Signaling contact
When there's a signal,
the relay conducts
current for a short time
e.g. in the case of
AUTOCAL:
Actuation of a second
device
Heating – gas path
Heating ready for
operation
For identification
purposes
Autocal check
Autocal difference too
great (function 24)
Combustion gas
Combustion gas valve
is open
Valve actuation
Combustion air
Combustion air valve is
open
Valve actuation
See also
List of faults (Page 6-6)
Motherboard pin assignments (Page 8-1)
Maintenance request and fault message (Page 6-3)
List of maintenance requests (Page 6-5)