Operation
5.10 Parameters
FIDAMAT 6-G
5-36
Operating instructions, 07/2006, A5E00478463
5.10.11
Measuring point switching (function 59)
59 Sample selection THC
M.pt. 1 Rel. 2 : 45: min
M.pt. 2 Rel. 1 2O min
M.pt. 3 Rel. 4 12 min
M.pt. 4 Rel. 3 8 min
M.pt. 5 Rel. 6 23 min
M.pt. 6 Rel. 5 6 min
M.p. switching on/off
Figure 5-40 Measuring point switching
With this function, you can allocate a maximum of six measuring points and switch them
cyclically and automatically.
The condition is that you parameterized the measuring point relay previously with function 71
"Relay allocation", which then actuates the corresponding solenoid valves.
Every measuring point relay is also allocated a time period, which you can enter in the
respective entry field. Values between 0 and 60 (minutes) are possible for this input.
Press the fifth softkey to activate/deactivate the measuring point switching..
You can allocate a signal relay to every measuring point relay. This allows there to be a
measuring point identifier separate from the measuring point relay. Configure the signal
relay, also with function 71.
See also
Relay allocation (function 71) (Page 5-40)
5.10.12
Logbook settings (function 60)
Figure 5-41 Parameterizing the logbook
With this function, you delete or block logbook entries.