Operation
5.11 Configuration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-55
5.11.12
Change codes (function 79)
79 Codes program.
THC
Code 1:111:
Code 2 222
Figure 5-58 Changing codes
With this function, you replace the factory-set codes ("111" for level 1, "222" for level 2) for
your own.
If you enter the value "000" for one code, the code block is lifted and you can have
immediate access to the corresponding operation level.
5.11.13
Device test (function 80)
8O Analyzer test
THC
Keyboard
Relay- and Binarytest
Analog test
Figure 5-59 Testing the device
With this function, you can test the function capability of the device.
Keyboard test
You can test all the keys on the control panel with the keyboard test. The five softkeys on the
right edge can be used to make the associated item appear/disappear. If a number key or
the sign key is pressed, this is displayed in the bottom line of the display.
After pressing the [INFO] key, a message is displayed in plain text. The [MEAS] and [ESC]
keys retain their "return" function.