Operation
5.10 Parameters
FIDAMAT 6-G
5-30
Operating instructions, 07/2006, A5E00478463
Description
Fct. no.
Remarks
Temperature recompensation of span
86
Temperature compensation
102
Only accessible for
service
Pressure compensation
104
Only accessible for
service
TC pre-compensation
111
Only accessible for
service
Linearization
113
Only accessible for
service
Besides the functions listed in the table "Functions switched on/off by function 52", function
52 can also be used to address other service functions. These are restricted to service
engineers and are only visible when the service code is input (code level 3).
5.10.5
Status messages (function 53)
53 Status messages THC
Display automat.
calibration [CAL]
Display stored
value [STO]
Display
limits [LIM]
Display
autorange [AR]
Display control
function [CTRL]
Figure 5-32 Status messages
With this function, you can parameterize the status message display in the status line of the
measured value display. You can display a maximum of three different status messages.
The message CODE ■/□ provides information about the current operating mode ("Coded
display mode"/"Decoded display mode") and is therefore always displayed.
The device activates the functional check (softkey 5 on the function screen) when it
determines that the measured value was influenced.
This is the case:
•
During the warm-up phase
•
After input of the code (also change into "operating mode")
•
During automatic calibration (Autocal) or remote calibration