Operation
5.1 General information for operation
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-3
Graphics symbols
■ = activated (ON state; also in status message in the status line)
□ = deactivated (OFF state; also in status message in the status line)
► = Entry in a submenu/subfunction
● = Triggering a function/subfunction (e.g. Start calibration...)
Device operating modes
Table 5-2
Operating modes of the FIDAMAT 6
Mode
Properties
Remarks
Coded display mode
•
Measured value display is
displayed
•
Protected submenus can
only be reached by entering
a code.
•
The current operating state
of the device (except for
"measure") is displayed in
the bottom line.
•
Functional check not active
The device only supplies
reliable measured values in this
mode, as far as it's in the
"measure" operating state. From
the operating mode, you get to
this mode by pressing the
[MEAS] button twice.
Decoded display mode
•
Measured value display is
displayed
•
The submenus protected by
the entered code are
accessible.
•
The current operating state
of the device (except for
"measure") is displayed in
the bottom line.
•
Functional check active
•
Measured value can be
influenced.
From the operating mode, you
get to this mode by pressing the
[MEAS] button once and
confirming or discarding the
made entries.
Operating mode
•
Menu or function is
displayed
•
Settings and input can be
entered
•
Functional check active
•
Measured value can be
influenced.
From "Coded display mode" you
get to this mode by entering the
code of the corresponding
operation level.
In this mode, you can
parameterize/calibrate the
device.