Installation and connection
3.4 Electrical connection
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
3-11
3.4.3.3
ELAN interface cable
Interface cable specification
Parameters
Value
Impedance
100 to 300 Ω for a measuring frequency of
> 100 kHz
Cable capacity
Typ. < 60 pF per meter
Core cross-section
> 0.22 mm2 , equivalent to AWG 23
Cable type
twisted pair, 1x2 conductors
Signal attenuation
max. 9 dB along the entire length of the
conductor segment
Shielding
Copper braiding shield or braided shield and foil
shield
Bus terminal resistors
For connection of bus terminal resistors, place a jumper from pin 3 to pin 7 in the first and
last connector of a bus line and from pin 8 to pin 9 (see figure "Pin assignment").
Note
For a conductor length longer than 500 m or high interference, we recommend that a
repeater be installed. For additional information, please refer to the ELAN interface
description.
Order numbers:
•
C79000-B5200-C176 German
•
C79000-B5276-C176 English