FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
4-1
Commissioning
4
4.1
Safety information
Warning
The following statements and information must absolutely be observed!
Danger
Explosion hazard
The gas analyzer FIDAMAT 6 is not permitted for measuring ignitable mixtures in normal
operation. If the device should be used for measuring flammable gases, which could form an
ignitable mixture with air or oxygen, special measures must be taken to protect against
explosion!
Danger
Electrical safety
Certain parts of this device are under dangerous voltage. Before switching on the device, the
housing must be sealed and grounded.
Neglecting to do so may result in death, body injury and / or material damage. Also, please
observe the chapter "Electric connection" for this.
Note
Flushing the housing
In accordance with the details in the “Technical data” section, there may be what is to be
regarded as a limited escape of flammable components from leaks in the sample path. With
the FIDAMAT 6 there is no need to flush the housing provided it has been ensured that there
is a natural exchange of air from the environment. This applies particularly when the unit is
installed in cabinets.