Operation
5.11 Configuration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-57
The device is always delivered in the ordered language. Usually, English is included as a
second language (if English is the first language, Spanish is set as the second language).
5.11.15
Interference gas correction (function 83)
83 Interf. correct. THC
Interference correct. 1
Interference correct. 2
Interference correct. 3
Interference correct. 4
Figure 5-62 Carrying out interference gas correction
Interference gas correction is usually not necessary with the FIDAMAT 6. Should you desire
interference gas correction in isolated cases, it is first to be distinguished what type of
interference gas influence there is.
Press the first softkey. The following function screen appears on the display.
By pressing the first softkey, now select the type of interference gas influence. The following
distinctions are possible:
•
No interference gas correction
•
Interference gas correction with constant interference gas influence
•
Interference gas correction with variable interference gas influence via analog input
•
Interference gas correction with variable interference gas influence via ELAN
The interference gas correction is deactivated for the duration of an calibration process (zero
or span). After calibration is complete and the device has jumped back to the "measure"
operating state, it is reactivated.