Operation
5.8 Calibration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-15
5.8.4
Setting setpoints (function 22)
22 Setpoints total THC
Setpoint for zero
: 5.4O: ppm C1
Setpoint for MR 1
8.5O ppm C1
Setpoint for MR 2
12.5O ppm C1
Setpoint for MR 3
6OO.5O ppm C1
Setpoint for MR 4
8245.5O ppm C1
Figure 5-11 Setting the setpoints
The function screen shows the setpoint input at total calibration. The fourth measuring range
is chosen as the leading measuring range.
With single calibration, there is no choice for the leading measuring range.
5.8.5
Total / single calibration (function 23)
Figure 5-12 Select calibration method
With this function, select between the total and single calibration of the measuring ranges.
Total calibration means that you adjust one "leading measuring range" and all other
measuring ranges are calibrated via the switching ratio.
If the total calibration, as shown on the function screen, is not activated, you must calibrate
every measuring range individually.