FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
7-1
Technical data
7
7.1
Technical data of the FIDAMAT 6 series
Table 7-1
Technical data - Structure, housing
Structure, housing
Degree of protection
IP 20 in accordance with EN 60529
Dimensions
See section "Dimension drawings and circuit
diagrams"
Weight
approx. 23 kg
Table 7-2
Technical data - electric features
Electric features
Power supply
AC 100 to 120 V (nominal consumption range
90 V to 132 V), 48 to 63 Hz or
AC 200 to 240 V (nominal consumption range
180 V to 264 V), 48 to 63 Hz or
Power consumption
approx. 150 VA in operation
approx. 350 VA in the warm-up phase
EMC interference immunity (electromagnetic
compatibility)
In accordance with the standard requirements of
NAMUR NE21 (08/98)
Electrical safety
In accordance with EN 61010-1
Overvoltage category II
Fuse values
100 to 120 V: 4.0 T/250
200 to 240 V: 2.5 T/250
Table 7-3
Technical data - gas consumption data
Gas consumption data
Combustion air
approx 350 ml/min
Hydrogen
approx. 20 ml/min
Sample gas
approx. 0.5 l/min
Zero gas / test gas
approx. 0.5 l/min