Commissioning
4.3 Commissioning and operation
FIDAMAT 6-G
4-4
Operating instructions, 07/2006, A5E00478463
Test pressure:
Operating pressure x 1.5 (operating pressure = 500 hPa)
Device electrically switched off:
Test pressure [hPa], rel.
Permissible pressure drop Δ p
[hPa] rel.
Sample gas path
750 +10
Δ p 10 +2
Supply gas input points
3.000 +100
Δ p 12 +5
Pressure equalization times: Test curve with respect to time
Note
After applying the test pressure, observe the start and end of measurement for determining
Δp.
Start of measurement x min
after applying test pressure
End of measurement y min after
applying test pressure
Sample gas path
2
3
Supply gas input points
1
2
4.3
Commissioning and operation
4.3.1
General information
Check
Before switching the unit on, check to see that combustion gas, combustion air, zero gas and
test gas are connected and the required pressures are present at the input points.
Switch on the power supply
Shortly after switch on, the measured value display appears on the display. At the upper
edge of the screen is the status line, at the lower edge, the bottom line. The current
operating state is displayed in the bottom line.
See also
Control panel and communication (Page 2-4)
Measured value display (Page 5-4)