Operator Manual
38
of 44
SPR2-310.620.30.02.02
System Description
Transport and parking position of the C-arm system
0
Prior to transport, the C-arm system must be set to the transport position.
Preparing the C-arm
◆
Release all brakes of the C-arm.
◆
Set the C-arm to the transport position shown in the drawing.
– Angulation 0°/Orbital position 0°.
◆
Lower the lifting column down to position 2.
(
→
Page 14)
◆
Move the horizontal carriage all the way back.
◆
Lock all brakes on the C-arm.
Disconnect the power
plug
◆
Switch off the ARCADIS Varic and wait for it to shut down.
◆
Pull the power plug out of the wall outlet.
Pull on the plug, not on the cable!
The ARCADIS Varic can be switched on again after approx. 5 s.
Please note that after finishing an examination, the ARCADIS Varic must be shut
down properly before it is disconnected from the power supply.