
Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02
21
of 34
Options
0
Dose measuring chamber (DAP meter)
0
Specifications
Work environment
Cassette holder
0
Technology
Ionization chamber
Active area
Diameter 90 mm, area 63.2 cm
2
DAP resolution
0.01
µ
Gym
2
/s equivalent to cGycm
2
/s
Maximum
measurable DAP
3000.00
µ
Gym
2
/s equivalent to cGycm
2
/s
Measuring inaccuracy
< 1% under constant pressure and temperature
Energy range
40 - 125 kV
±
6%
Attenuation equivalent
(Inherent filtration)
< 0.5 mm Al equivalent
Dose rate linearity
better than
±
2%
Measuring readiness
10 seconds after power on
Temperature
+10 °C to +70 °C
Relative humidity
15% to 75% (without condensation)
Barometric pressure
700 hPa to 1060 hPa
Format
24 cm x 30 cm (10 inches x 12 inches)
Format with grid
Pb r17 N70, f
o
85
Содержание ARCADIS Varic
Страница 2: ......
Страница 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Страница 6: ......
Страница 8: ......
Страница 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Страница 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Страница 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Страница 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Страница 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Страница 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Страница 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Страница 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...