Safety
ARCADIS Varic
SPR2-310.620.30.02.02
13
of 28
Overload protection
0
Prolonged continuous radiation at maximum tube load is permissible in fluoros-
copy mode. However, this can cause the X-ray tube assembly to heat up. For this
reason, the X-ray tube assembly has a thermal monitor. If necessary, power is
reduced in all operating modes, in SUB/Roadmap with the next new scene/mask.
If the temperature rises to
≥
50 °C,
❏
the temperature indicator on the control panel of the C-arm system lights up.
❏
the selected characteristic curve is switched to S1 at the end of radiation in
fluoro and pulsed fluoro.
If the temperature rises to
≥
60 °C,
❏
the selected characteristic curve is switched to S1 during radiation in fluoro
and pulsed fluoro.
If the temperature rises to
≥
70 °C,
❏
the temperature indicator on the control console of the C-arm system flashes.
❏
radiation is aborted and cannot be released again.
If the temperature falls below 50 °C,
❏
the previously deselected curve is automatically reselected.
Содержание ARCADIS Varic
Страница 2: ......
Страница 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Страница 6: ......
Страница 8: ......
Страница 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Страница 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Страница 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Страница 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Страница 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Страница 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Страница 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Страница 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...