Maintenance
ARCADIS Varic
SPR2-310.620.30.02.02
7
of 16
Updating virus protection
0
Virus protection should be updated every 2-3 days to make sure the
ARCADIS Varic has the latest virus definitions installed. The update process
takes a few minutes and should be performed during a break or outside of exam-
ination hours.
Calling up the dialog
window
To update the virus protection, open the
SoftwareDistribution
dialog window in
the
Local Service
window.
◆
Select
Options > Customer Service > Local Service
in the main menu of the
task card.
– The
Service Software
dialog box is displayed.
◆
Delete the default code in the left window of the entry area
Please enter
password
and click
OK
.
– The start window for service settings is displayed.
◆
Click
SoftwareDistribution
.
– The
SoftwareDistribution
window appears.
Содержание ARCADIS Varic
Страница 2: ......
Страница 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Страница 6: ......
Страница 8: ......
Страница 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Страница 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Страница 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Страница 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Страница 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Страница 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Страница 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Страница 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...