Operator Manual
8
of 24
SPR2-310.620.30.02.02
Options
Multi-function footswitch (option)
0
MODE
Warning
When the C-arm is rotated by 180° and lowered to maximum, contact between
the I.I. and the footswitch could result.
Unintentional release of radiation!
u
Please make sure that the footswitch is not located underneath the I.I.
Caution
If the C-arm is lowered all the way and the bracket is folded up, there may be
contact between the multifunctional footswitch and image intensifier or tube
assembly housing.
Damage to the multifunctional footswitch and image intensifier or tube
assembly housing!
u
Please make sure that the footswitch is not located underneath the I.I.