Maintenance
ARCADIS Varic
SPR2-310.620.30.02.02
5
of 16
Checks prior to special examinations (e.g. of the open
heart and skull)
0
◆
Make sure that there is an additional conductive connection between the
C-arm system and a point of potential equalization, e.g. the patient table.
Dose and consistency test
0
The ARCADIS Varic system features automatic object position detection.
Proceed as follows:
◆
Place the object or filter into the beam path.
◆
Release fluoroscopy in automatic mode for approx. 10 s.
◆
Press the ADR stop button on the control panel.
◆
Attach the dose measuring chamber to the image intensifier input.
◆
Release fluoroscopy again to perform the dose measurement.