Options
ARCADIS Varic
SPR2-310.620.30.02.02
23
of 24
Deleting the diaphragm positions
0
◆
If you wish to delete the stored diaphragm positions again, press one of the
diaphragm keys and the diaphragm will open completely when the cassette is
inserted.
Setting exposure data
0
If a cassette is loaded after the system is restarted, only dashes will be displayed
in the kV and mAs fields. You cannot release an exposure since this can only be
done after setting the required kV and mAs values.
◆
Now press one of the kV or mAs +/- buttons.
– The preference values 40 kV and 5 mAs will be displayed. You can then
change these values.
◆
Now press one of the kV or mAs +/- buttons until the required values are set.
– If you perform fluoroscopy in the meantime, the kV and mAs values used
for the last cassette exposure will remain set when reloading the cassette.
Example of kV and mAs values:
Display of kV values
Display of mAs values
kV
mAs
mA
Содержание ARCADIS Varic
Страница 2: ......
Страница 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Страница 6: ......
Страница 8: ......
Страница 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Страница 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Страница 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Страница 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Страница 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Страница 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Страница 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Страница 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...