Examination
ARCADIS Varic
SPR2-310.620.30.02.02
15
of 52
Exam set
0
Three different dose programs are available for each selectable body region.
These programs are contained in the operating mode settings.
When a body region is selected, the
Standard Dose
program is activated:
Modifying the dose
Usually a dose rate deviating from the standard setting is reasonable only if you
have to adapt the dose to the particular anatomical conditions of the patient
(e.g. very slim or obese patients).
◆
Use the selection list to select another exam set (dose program), if necessary.
– With
Reduced Dose
you set the minimum dose rate which provides a sat-
isfactory image quality for the selected body region.
If you change the exam set, the fluoroscopy mode is automatically set.
Even if a different LUT is stored with the new exam set, the display of the current
image (LIH) or of the subtracted image does not change (the current LUT is main-
tained).
Содержание ARCADIS Varic
Страница 2: ......
Страница 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Страница 6: ......
Страница 8: ......
Страница 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Страница 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Страница 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Страница 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Страница 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Страница 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Страница 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Страница 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...