. Application Functions
Notes
• Data may be output when a container is loaded, however this is not an
error.
• On the premise of weighing animated objects, the stability detection band is
automatically extended in the Animal Weighing mode. Reproducibility of the
measurement data is slightly less than with other modes. If a larger stability
detection band is selected with the menu (See
8.3.1
), stability detection is more
readily made and the output of measurement result usually becomes faster.
However, the accuracy may become poorer. When changing stability detection
band, select according to the specific measurement’s requirements on
accuracy, observing the reproducibility on the test run.
• See
8.3
for details of stability detection.
• If the balance is slow to automatically return to the zero display, set a
larger Zero Range value.
• See
10.5
for details of Zero Range.
• When Menu Lock (
4.5.3
) is activated, operational condition of Animal
Weighing is also locked.
• When Menu reset (
4.5.2
) is performed, operational condition setting of
Animal Weighing is also returned to the default (Cond 1).
• Animal Weighing mode setting can be cleared without menu operation, by
pressing the [POWER] key for about three seconds.
• Pretaring Value (
10.7
) function cannot be used with Animal Weighing
mode.
• Small animal bucket set (option, see
A-4.
) is recommended for weighing
small animals such as mice.
• Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print (
10.3
), Auto Zero (
10.4
), Peak Hold (
11.3
), Interval Timer
(
11.4
), Add-on mode (
11.5
), Animal Weighing mode (11.6), Formulation
mode (
11.7
). If a function is set via the menu while another function is
already set, the original function will be automatically cancelled.
• When minimum display has been switched to 10d (See
6.2
), the optimized
Zero Range for 1d display may cause failure in the operation (e.g. Data is
not automatically output as animal is placed). Select optimal Zero Range
value again when minimum display has been changed (See
10.5
for
details).
The UP Series animal weighing mode (8.2) cannot be used. Change to UW/
UX Series automatic mode.
• UP Series automatic mode (10.2) cannot be used. Change to UW/UX
Series automatic mode.