![Shimadzu UP1023X Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/shimadzu/up1023x/up1023x_instruction-manual_1256959102.webp)
87
12
. Connections and Communications with External Equipment
2
4
1
3
5
6
7
11
13
12
8
9
10
14
12.5
Communication Setting
12.5.1
Overview
This menu is used to set the specifications for communication between the balance and a personal
computer or electronic printer.
This menu affects both the RS-232C and DATA I/O at the same time. For the instrument to be connected
to the DATA I/O connector such as an electronic printer, select the communication setting of the balance
to the default settings, which are menu items [H-tm]
76
, [F-Eb]
77
, [b-1200]
83
, [P-no]
89
, [S-1]
92
,
[d-Cr]
94
.
12.5.2
Handshaking
Determine the specifications for handshaking.
To have software handshaking not performed, select menu item
73
[H-oFF]
To have software handshaking performed as the following,
select menu item
74
[H-SoFt]
After the balance receives X-OFF (13H), the balance output is retained.
After the balance receives X-ON (11H), the balance output is initiated.
To have hardware handshaking performed as the following,
select menu item
75
[H-HArd]
When DTR is OFF, the output from the balance is retained.
When DTR is ON, the output from the balance is initiated.
To have timed hardware handshaking performed,
select menu item
76
[H-tm]
Notes
• Handshaking determines whether the peripheral equipment can receive
communication data from the balance. This function does not relay the
status of the balance to the peripheral equipment.
• The balance is able to receive as long as there is space in the receiving
buffer of the balance. This function operates once “oFF” is displayed,
operation in other states is not guaranteed.
• When the balance output is retained by handshaking, the display of the
balance is locked. Determine the specifications for handshaking.