. Application Functions
Notes
Up to four decimal places are displayed for density or specific gravity.
Depending on the measurement conditions, not all digits are stable even
the balance is working normally. The [UNIT] key can also be used to
change the minimum display.
How to change the minimum display in density (specific gravity)
display
When the result is displayed, press and keep holding down the [UNIT] key
for about three seconds. This changes the minimum display to 10d (d:
original minimum display). Every time the same key operation is made, the
minimum display changes and cycles in the order of “1d
→
10d
→
100d
→
1000d
→
1d
→
”.
• When loading the reference weight on the pan in the liquid, ensure that the
entire weight is immersed in the liquid.
• The dimensions of the below-weigh hook are shown in
A-5
.
• UP Series automatic mode (10.2) cannot be used. Change to UW/UX
Series automatic mode.