70
11
. Application Functions
11.6
Animal Weighing mode
Not applicable to a verified balance as a legal measuring instrument in using region
Designed for weighing live animals. Select menu item
53
to activate Animal Weighing mode. Also,
optimize the operational condition (See the following), “Stability Detection Band (
8.3.1
)” and “Timing of
Stability Mark Illumination and Data Output (
8.3.2
)” depending on the size and movement of the weighed
animal. If you weigh rats or mice, refer to the below table showing the typical settings to obtain optimal
accuracy and efficiency.
Recommended Settings for Rat and Mouse
Animal
Operational
condition
(See below)
Stability detection band
(See
8.3.1
for details)
Timing of Stability Mark Illumination
and
Data Output (See
8.3.2
for details)
Rat
Cond3
4 counts (menu item
29
)
ES-FASt (menu item
21a
)
Mouse
Cond1 or 2
1 count
(menu item
27
= default setting)
ES-FASt (menu item
21a
)
Operational condition selection according to the level of animal’s movement
Depending on the level of animal’s movement, operational condition (response and stability) can be
optimized within Animal Weighing mode. At zero display in Animal Weighing mode, operational conditions
can be switched using only the [UNIT] key. Select the optimal operational condition by observing the
response and stability of measurement.
(3 operation modes)
Operational mode
Level of animal’s movement
Remarks
Cond 1
Calm
Cond 2
Moderate
Not suitable for weighing animal lighter than 50g.
Cond 3
Active
Not suitable for weighing animal lighter than 100g.
Cond 2 and Cond 3 are not suitable for weighing animals lighter than the respectively specified in the
above. When lighter animals are weighed with these settings, the zero return after removing the animal
may be very slow.
(Switching operational conditions)
Every time the [UNIT] key is pressed at the zero display in Animal Weighing mode, the operational
condition will be changed in the following order: The operation mode (Example: [Cond 1]) will be displayed
temporarily when switching modes. Thereafter, the confirmation of the operational condition setting can be
made observing the position of the symbol in the right end of the display.