10
14
UPY Series
[Span Calibration Using External Weights]
Note
Not applicable to a verified balance as a legal measuring instrument in using
region.
1
Verify that the balance is in gram-display and
unload the sample from the pan.
2
Press the [CAL] key once. “E-CAL” is displayed.
3
Press the [O/T] key.
The value of the correct calibration weight to be
loaded is displayed and blinks.
Change the weight value to be used
Press the [CAL] key to change the weight value.
Adjust the value using the [UNIT] key and [PRINT]
key then press the [O/T] key. This will set the
modified weight value. Press the [POWER] key to
interrupt modification.
Refer to 5.4 for numerical value input methods.
The range of weights that can be used for
sensitivity calibration is designated separately for
each model. Refer to "External weight calibration
ranges" in "A1. Specifications".
Note that attempting to input a value outside of
the range will cause an error.
4
Load the indicated calibration weight and press
the [O/T] key.
(Example)
UP0000x
3