39
7
. Calibration
2
4
1
3
5
6
7
11
13
12
8
9
10
14
7.3.3
Clock-CAL Fully-automatic Calibration (UPX Series only)
In the UPX Series, automatic span calibration of the balance can be carried out at a set time.
It is possible to set up to three specific times for Clock-CAL (“tCALt1”, “tCALt2”, and “tCALt3”). Use the 24
hour system to set menu items
7
,
8
and
9
.
• Setting to “00:00” releases the function.
Notes
To execute Clock-CAL, all of the following conditions must be satisfied at the set
time. If these conditions are not satisfied within one minute, the automatic span
calibration is not executed and that cycle is skipped.
• There must be nothing on the pan.
• The balance must be in mass display or the standby mode.
• If in mass display mode, the display must be within a 30 count
• The balance must be stable. (The stability symbol must be illuminated
during mass display.)
• Load on the pan should be near zero.
• No other automatic span calibration (PSC or timer CAL) must have
occurred during the same time
Example: Setting “tCAL t1” to twelve noon.
1
Select menu item
7
2
Set the desired time.(Refer to
4.4
for numerical
input.)
3
Press the [O/T] key. The set time will be stored.
4
Proceed to next time setting by the [MENU] key,
or return by the [POWER] key.
Skipping Clock-CAL
If Clock-CAL starts while the balance is in use, press the [POWER] key to abort that cycle.
(Example)
Содержание UP1023X
Страница 3: ...No text...
Страница 124: ...109 2 4 1 3 5 6 7 11 13 12 8 9 10 14 Appendices...