![Shimadzu UP1023X Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/shimadzu/up1023x/up1023x_instruction-manual_1256959075.webp)
60
11
. Application Functions
After entering the value, set using the [O/T] key,
then press the [POWER] key two times to return
to mass display.
2
Remove the below-weigh hook cap from the
bottom of the balance to expose the below-
weigh hook.
3
Hook the hanging pan, and then immerse the
hanging pan in the tank filled with the liquid.
4
When the [UNIT] key is pressed a number of
times from the mass display, [Air
▼
gd] appears for
about two seconds. Afterward, the display
switches to the [
▼
gd
▶
] display. This is the weight
measurement in air mode. During weight
measurement in air, “ ” turns on in the upper-
right of the display.
5
Press the [O/T] key.
6
Place the items to be measured on the pan.
7
When the stability mark appears, press the [CAL]
key. This measures the weight of the item being
measured in air.
8
The display shows [wAtEr
▼
gd] for about two
seconds. Afterward, the display switches to the
[
▼
gd
▶
] display. This is the weight measurement
in water mode. During weight measurement in
water, “ ” turns on in the lower-right of the
display.
9
Place the items to be measured on the
immersed hanging pan. The display shows the
weight value in water. Press the [CAL] key to
show the specific gravity value in the set specific
gravity value display mode. Press the [POWER]
key to return to
8
. If air bubbles adhered to the
item being measured and the result is a specific
gravity value that differs from the expected value,
you can recalculate the specific gravity value by
retrying the weight measurement in water after
removing the air bubbles from the item.
(Example)
(Example)
(Example)
Hold display mode example
Содержание UP1023X
Страница 3: ...No text...