When the [UNIT] key is pressed a number of
times from the mass display, [Airgd] appears for
about two seconds. Afterward, the display
switches to the [gd
▶
] display. This is the sinker
weight measurement in air mode. During weight
measurements in air, “ ” turns on in the
upperright of the display.
5
Press the [O/T] key.
6
Place the sinker on the balance pan.
7
When the stability mark appears, press the [CAL]
key. This measures the weight of the sinker in air.
8
[wAtEr gd] appears for about two seconds.
Afterward, the display switches to the [gd
▶
]
display. This is the sinker weight measurements
in liquid, “ ” turns on in the lower-right of the
display.
9
Place the sinker in the immersed hanging pan.
The display shows the weight value in liquid.
Press the [CAL] key to show the liquid density in
the set specific gravity value display mode. Press
the [POWER] key to return to
8
. If air bubbles
adhered to the sinker and the result is a density
that differs from the expected value, you can
recalculate the density by retrying the weight
measurement in liquid.
10
To make the next measurement, press [CAL],
and begin again at Step
5
.
(Example)
(Example)
Hold display mode examlpe