Troubleshooting
Symptom
Probable cause(s)
Countermeasure
Nothing is displayed.
• The AC adapter is disconnected.
• The breaker of the room is off.
• The voltage is wrong.
Check the power and connect AC adapter
correctly.
“OL” or “-OL” is displayed.
Transportation screws haven’t been
loosened.(UPX only)
Turn them anti-clockwise until they stop.
The pan is displaced.
Place the pan properly.
Pan supporter caps are not installed.
Install pan supporter caps.
The load on the pan is too large.
Use balance within its capacity.
Display does not change
when pan is loaded.
Pan is displaced.
Place the pan properly.
The display fluctuates and
it is difficult to get the
stability symbol to display.
Affected by vibration or air flow.
Install the balance at a proper site.
Try changing environment settings.
Protective in-use cover touches the pan.
Adhere the cover firm to the balance body.
The weighed result is not
accurate.
Span calibration has not been done.
Calibrate it properly.
Balance was not zeroed before
measurement.
Press the [O/T] key to zero the mass display
before measuring.
Does not display the unit
desired to use.
The unit has not been set up
Set it up in advance.
Menu item selection is
rejected.
Menu Lock is ON.
Release Menu Lock. (See
4.5.3
)
An error code is shown.
↑
See
“ Error Display”
(
14.2
)
14.4
LCD (Liquid Crystal Display) Check
The display can be configured to standby in the whole lighting display mode when pressing the [POWER]
key in the power supply standby state to switch to mass display mode. If set to [E8-StoP], numbers will be
displayed like an electronic scoreboard when powering on for easier verification.
Mode for temporarily stopping in whole lighting display mode for standby
select the menu item
37
[E8-StoP]
When the display holds, press the [O/T] key to proceed to the
mass display.
Mode for automatically switching to mass display after approximately 0.5 seconds of whole
lighting display
select the menu item
38
[E8-Cont]
If all items are not displayed in the whole lighting state as shown in the figure in 1.3, contact the service
company, etc.
• Refer to Chapter 4 for menu setting details.