36
7
. Calibration
7.2.4
Calibration Check Using External Weights
Not applicable to a verified balance as a legal measuring instrument in using region
Use of a standard calibration weight to check and display the degree of balance sensitivity drift.
1
Verify that the balance is in mass display and that
the pan is empty.
2
Press the [CAL] key once to display “E-tESt”.
If “E-tESt” is not displayed, select menu item
4
.
3
Press the [O/T] key.
The value of the correct calibration weight to be
loaded is displayed and blinks.
Changing the Calibration Weight to be Used
Pressing the [CAL] key allows changes to the weight
value. Modify the value using the [UNIT] key and
[PRINT] key, then press the [O/T] key. Refer to
4.4
for
how to make numerical input.
To interrupt modification, press the [POWER] key.
Calibration range with external weights is designated
to each model. Refer to
“1. Specifications”
for
calibration range. Attempt of inputting an invalid
calibration weight value causes an error message.
4
Load the indicated calibration weight and press
the [O/T] key. The zero display blinks.
5
Unload the weight from the pan and press the
[O/T] key.
The display changes to the “d xxx” display.
6
Hereafter the procedure is the same as item
4
in
7.2.2. Calibration Check Using the Built-in
Weight.
(Example)