66
11
. Application Functions
Notes
• Press the [POWER] key in the peak detection standby state to initiate the
power standby state.
• Press the [POWER] key during detection of the peak to return to the peak
detection standby state.
• Peak Hold setting can be cleared without menu operation, by pressing
the [POWER] key for about three seconds.
• Polarity of the peak value displayed is “polarity of the displayed value of
the first change (by five times or more of Zero Range) from the display
within Zero Range.”
• Usually the peak value is easily measured by selecting menu item
23
[EA-PoUr](Pouring mode). Depending on the weighting conditions and the
sample type, this setting may not always be suitable.
• Pretaring Value (
10.7
) function cannot be used with Peak Hold function.
• Refer to 8.3 for stability detection details, and 10.5 for zero range details.
• Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(
10.3
), Auto Zero(
10.4
), Peak Hold(
11.3
), Interval
Timer(
11.4
), Add-on mode(
11.5
), Animal Weighing mode(
11.6
), Formulation
mode(
11.7
). If a function is set via the menu while another function is
already set, the original function will be automatically cancelled.
• When minimum display has been switched to 10d (See
6.2
), the optimized
Zero Range for 1d display may cause failure in the operation. Select
optimal Zero Range value again when minimum display has been
changed (See
10.5
for details).