![Shimadzu UP1023X Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/shimadzu/up1023x/up1023x_instruction-manual_1256959047.webp)
32
7
. Calibration
7.2
Calibration Execution
Notes
• Setting before shipment is as the following:
UPX Series: Span calibration using the built-in weight ([i-CAL])
UPY Series: Span calibration using external weights ([E-CAL])
The type of calibration can be changed (See 7.3).
• Calibration will not be performed when the weight on the pan is not near
zero, or the balance is not stable.
7.2.1
Span Calibration Using the Built-in Weight (UPX Series Only)
The balance is adjusted using the built-in calibration weight.
1
Verify that the balance is in mass display and that
the pan is empty.
2
Press the [CAL] key once. “i-CAL” is displayed.
If this indication is not displayed, select menu
1
.
Press and hold the [POWER] key for three
seconds after selecting to return to the g display.
3
Press the [O/T] key
[ i-CAL3 ] ~ [ i-CAL 1], [ SEt ], [ CAL End] will be
displayed in that order. Then the mass display
will appear indicating the completion of span
calibration.
…