Remote Control Commands
R&S
®
SMF100A
557
Operating Manual 1167.2319.02 ─ 12
Example:
SWE:DWEL 12 ms
sets a dwell time of 12 ms for a frequency sweep at the RF out-
put.
Manual operation:
See
" Dwell Time - Frequency Sweep "
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Starts an RF frequency sweep cycle manually.
The command is only effective in single mode.
Example:
TRIG:FSW:SOUR SING
SOUR:SWE:FREQ:MODE AUT
SWE:FREQ:EXEC
triggers a frequency sweep at the RF output.
Usage:
Event
Manual operation:
See
" Execute Single Sweep - Frequency Sweep "
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer:ACTive
<Active>
The command activates the selected marker. The active marker is output with a higher
voltage than all other markers.
Parameters:
<Active>
NONE | M01 | M02 | M03 | M04 | M05 | M06 | M07 | M08 | M09 |
M10
*RST:
NONE
Example:
SOUR:SWE:FREQ:MARK:ACT M05
sets marker 5 to be the active marker.
Manual operation:
See
" Active Marker - RF Frequency Marker "
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer:AMPLitude
<Amplitude>
Sets the amplitude marker attenuation.This value is valid for all markers.
Correlation:
The output level reduction is applied to a marker with command
SOUR:SWE:MARK<[1]...10>:PST ON
.
Parameters:
<Amplitude>
float
Range:
0 to 3
Increment: 0.1
*RST:
1
Example:
SOUR:SWE:FREQ:MARK:AMPL 1
the output level is reduced by 1 dB on reaching the mark.
Manual operation:
See
" Amplitude - RF Frequency Marker "
SOURce Subsystem