Remote Control Commands
R&S
®
SMF100A
543
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:PULM:TRAin:ONTime
<OnTime>
Fills the On-time part of the selected file with data.
Parameters:
<OnTime>
Ontime#1{, Ontime#2, ...} | binary block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data.
When block data is transferred, 8 (4) bytes are always interpre-
ted as a floating-point number with double accuracy (see the
command
FORMat:DATA
).
The maximum length is 2047 values.
Example:
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:SEL ‘P_INCR'
selects
P_INCR
for editing.
P_INCR
is created if it does not yet
exist.
PULM:TRA:ONT 10ns,30ns,40ns,...
specifies the on-time values in
P_INCR
. If the list already con-
tains data, it is overwritten.
Options:
R&S SMF-K27 (Pulse Train)
Manual operation:
See
[:SOURce<hw>]:PULM:TRAin:ONTime:POINts?
Queries the length (in points) of the ontime component of the selected list.
Return values:
<Points>
integer
Range:
0 to 2047
*RST:
0
Example:
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:SEL ‘P_INCR'
selects
P_INCR
for editing.
P_INCR
is created if it does not yet
exist.
PULM:TRA:ONT:POIN?
queries the number of frequency values in
P_INCR
Response: 7
P_INCR
has 7 ontime entries.
Usage:
Query only
Options:
R&S SMF-K27 (Pulse Train)
SOURce Subsystem