Instrument Function
R&S
®
SMF100A
157
Operating Manual 1167.2319.02 ─ 12
Figure 5-3: Synchronizing instruments by means of an externally applied reference signal having
10 MHz
Settings:
–
–
Set the additionally provided parameters, as for example the synchronization band-
width according to the requirements of the application.
Input and output connectors of the reference frequency
The appropriate connectors are located at the rear panel, see
To use the EFC feature, apply the signal at the EFC input connector at the rear of the
instrument, see
5.3.4.1
Reference Oscillator Settings
To access the settings dialog for configuring the reference signal, perform one of the
following:
●
In the block diagram, select "RF Frequency > config... > RF Frequency > Refer-
ence Oscillator"
●
Press the MENU key and select "RF Frequency > RF Frequency > Reference
Oscillator"
●
Press the SETUP key and select "Setup > System > Reference Oscillator"
RF Frequency Block