Instrument Function
R&S
®
SMF100A
164
Operating Manual 1167.2319.02 ─ 12
"Extern Step"
Generates the sweep signal step-by-step, manually triggered.
To trigger a sweep step, apply an external trigger signal. The step
width corresponds to the step width set for the rotary knob.
Example:
SOUR:SWE:FREQ:MODE STEP
SOUR:SWE:FREQ:SPAC LIN
SOUR:SWE:FREQ:STEP:LIN 1MHz
TRIG:FSW:SOUR EXT
SOUR:FREQ:MODE SWE
(External trigger)
"Extern Start/Stop"
Generates a continuously repeating sweep signal that is started,
stopped and restarted by subseqeunt external trigger events. The
sweep steps are performed automatically, controlled by the dwell
time.
Refer to the description of the rear panel for information on the con-
nectors for external trigger signal input (see
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR EAUT
SOUR:FREQ:MODE SWE
(External trigger)
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
Specific Sweep Parameters
The sweep parameters described in this topic are only available for particular emula-
tion languages.
For details, see
"Remote Emulation with the R&S
SMF microwave signal gen-
erator".
Mode ← Specific Sweep Parameters
Selects the RF frequency sweep mode.
("Auto" = timer controlled, "Manual" = manually controlled)
Repeat ← Specific Sweep Parameters
Sets the repetition mode.
("On" = continuously; "Off" = once)
Start Trigger ← Specific Sweep Parameters
Selects the mode to initiate a sweep.
("Auto" = starts immediately; "Single" = starts with a command (Bus); "External" =
starts with an external trigger signal)