Remote Control Basics
R&S
®
SMF100A
339
Operating Manual 1167.2319.02 ─ 12
3. Send the overlapped command with
*OPC
.
4. Wait for a service request.
The service request indicates that the overlapped command has finished.
*OPC? with a service request
1. Set bit no. 4 in the SRE:
*SRE 16
to enable MAV service request.
2. Send the overlapped command with
*OPC?
.
3. Wait for a service request.
The service request indicates that the overlapped command has finished.
Event status register (ESE)
1. Set the OPC mask bit (bit no. 0) in the ESE:
*ESE 1
2. Send the overlapped command without
*OPC
,
*OPC?
or
*WAI
.
3. Poll the operation complete state periodically (with a timer) using the sequence:
*OPC; *ESR?
A return value (LSB) of 1 indicates that the overlapped command has finished.
6.5 Status Reporting System
The status reporting system stores all information on the current operating state of the
instrument, and on errors which have occurred. This information is stored in the status
registers and in the error queue.
You can query both with the commands of the
6.5.1 Hierarchy of the Status Registers
shows the hierarchical structure of information in the status registers
(ascending from left to right).
Status Reporting System