![Rohde & Schwarz SMF-B1 Скачать руководство пользователя страница 180](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740180.webp)
Instrument Function
R&S
®
SMF100A
165
Operating Manual 1167.2319.02 ─ 12
Point Trigger ← Specific Sweep Parameters
Selects the mode to trigger the progress of each step.
("Auto" = immediately; "Single" = remotely (Bus); "External" = external signal)
Execute Single Sweep - Frequency Sweep
Starts a sweep manually. This trigger button is displayed in "Single" mode.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>[:SWEep][:IMMediate]
Reset Sweep - Frequency Sweep
Resets the sweep.
With the next trigger event, the sweep starts with at the initial value.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL]
Start Freq - Frequency Sweep
Sets the start frequency.
Remote command:
Stop Freq - Frequency Sweep
Sets the stop frequency.
Remote command:
Center Freq - Frequency Sweep
Sets the center frequency.
Remote command:
[:SOURce<hw>]:FREQuency:CENTer
Span - Frequency Sweep
Sets the span.
Remote command:
Current Freq - Frequency Sweep
Displays the current frequency.
In sweep "Step" mode, the parameter is editable and you can enter frequency for the
next step.
Remote command:
[:SOURce<hw>]:FREQuency:MANual
Spacing - Frequency Sweep
Selects the mode for the calculation of the frequency sweep intervals.
RF Frequency Block