Remote Control Commands
R&S
®
SMF100A
488
Operating Manual 1167.2319.02 ─ 12
●
logarithmic sweeps and f
STARt
< f
STOP
freq_points = ((log f
STOP
- log f
STARt
) / log step_log) + 1
To determine the logarithmic step size, use the command
.
If you change the number of sweep points, the step size changes accordingly. The
sweep range remains the same.
Each sweep spacing mode has assigned the
POINts
setting separately. Thus, the
command refers always to the particular set mode, see
.
Note:
The number of frequency sweep points displayed in the dialog deviates approxi-
mately by 1 from the frequency sweep points set with this remote control command.
It is up to the different clculation methods:
●
the remote control command determines the number of steps.
●
in the dialog, the instrument displays the number of transitions according to:
Count = ( f
STOP
- f
STARt
) / step_log
Since
Count
is an integer, the result may be rounded.
Parameters:
<Points>
integer
Range:
2...max
Example:
LFO:FREQ:STAR
sets the start frequency to 2 kHz.
LFO:FREQ:STOP
sets the stop frequency to 20 kHz
LFO:SWE:SPAC LIN
sets linear sweep spacing.
LFO:SWE:POIN 11
sets 11 sweep steps for linear sweep spacing. The sweep step
width (
STEP
) is automatically set to 2 kHz.
Options:
R&S SMF-B20
Manual operation:
See
" Count - LF Frequency Sweep "
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RETRace
<State>
Activates that the signal changes to the start frequency value while it is waiting for the
next trigger event.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
SOURce Subsystem