Instrument Function
R&S
®
SMF100A
229
Operating Manual 1167.2319.02 ─ 12
Remote command:
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge[:STATe]
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:SFRequency:STATe
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:SFRequency
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:SFRequency:STATe
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:SFRequency
Min Frequency - Power Analysis
Power versus frequency measurement only with active "Use Separate Frequency".
Sets the minimum frequency of the measurement.
Remote command:
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STARt
Max Frequency - Power Analysis
Power versus frequency measurement only with active "Use Separate Frequency"
only.
Sets the maximum frequency of the measurement.
Remote command:
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STOP
Use as X Scale - Power Analysis
Measurement Mode Frequency only.
Activates the use of the separate frequency min and max values for the scaling of the
x-axis. Thus, the trace for this sensor is visible in the diagram , especially for frequency
ranges that differ substantially from the generator settings.
If more than one sensor is active, which use separate frequencies, the option is only
available for one sensor. To indicate the traces of the other sensors, use function "Map
to X Scale".
Remote command:
n.a.
Map to X Scale - Power Analysis
Mode Frequency and active "Use Separate Frequency" only.
Maps the trace of a sensor that uses separate frequency to the current scaling of the
diagram. Usually the scale is determined by the set frequency range of the generator.
If more than one sensor is active, which use separate frequencies, the scale can also
be determined by the separate frequency range of one of the other sensors.
Remote command:
n.a.
RF Level