Remote Control Commands
R&S
®
SMF100A
406
Operating Manual 1167.2319.02 ─ 12
Options:
Option R&S SMF-K28
Manual operation:
See
:SENSe[:POWer]:SWEep:FREQuency:STARt
<Start>
Sets the start frequency for the frequency mode.
Parameters:
<Start>
float
Range:
0 to 1E12
*RST:
1E6
Example:
SENS:SWE:FREQ:STAR 2.0GHZ
'sets a sweep start at 2 GHz.
Options:
Option R&S SMF-K28
Manual operation:
See
:SENSe[:POWer]:SWEep:FREQuency:STEPs
<Steps>
Sets the number of measurement steps for the frequency mode.
Parameters:
<Steps>
integer
Range:
10 to 1000
*RST:
200
Example:
SENS:SWE:FREQ:STEP 500
sets 500 steps
Options:
Option R&S SMF-K28
Manual operation:
See
:SENSe[:POWer]:SWEep:FREQuency:STOP
<Stop>
Sets the stop frequency for the frequency mode.
Parameters:
<Stop>
float
Range:
0 to 1E12
*RST:
22GHZ
Example:
SENS:SWE:FREQ:STOP 20.0GHZ
sets the sweep stop to 20 GHz
Options:
Option R&S SMF-K28
Manual operation:
See
Power Sensor Measurement Subsystems