Remote Control Commands
R&S
®
SMF100A
429
Operating Manual 1167.2319.02 ─ 12
Parameters:
<SFrequency>
float
Range:
0 to 1E12
Increment: 1
*RST:
1E6
Example:
SENS1:SWE:TIME:SENS:SFR 2GHz
the measurement is performed at 2 GHz
Options:
Option R&S SMF-K28
Manual operation:
See
" Use Separate Frequency- Power Analysis "
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:SFRequency:STATe
<State>
Activates the use of a different frequency for the power measurement.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SENS1:SWE:TIME:SENS:SFR:STAT ON
activates the use of a separate frequency than the generator fre-
quency for power analysis
Options:
Option R&S SMF-K28
Manual operation:
See
" Use Separate Frequency- Power Analysis "
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:AUTO
<Auto>
Sets the trigger level, the hysteresis and the dropout time to default values.
Parameters:
<Auto>
ONCE
*RST:
---
Example:
SENS1:SWE:TIME:SENS:TRIG:AUTO ONCE
the trigger level is automatically determined
Options:
Option R&S SMF-K28
Manual operation:
See
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:DTIMe
<DTime>
Determines the minimum time for which the signal must be below (above) the power
level defined by level and hysteresis before triggering can occur again.
Parameters:
<DTime>
float
Range:
0 to 10
*RST:
200E-9
Power Sensor Measurement Subsystems