Remote Control Commands
R&S
®
SMF100A
404
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Offset>
float
Range:
-100 to 100
Increment: 0.01
*RST:
0
Example:
SENS2:SWE:FREQ:OFFS -3dB
defines a level offset of -3 dB.
SENS2:SWE:FREQ:OFFS:STAT ON
activates the specified level offset.
Options:
Option R&S SMF-K28
Manual operation:
See
" Level Offset - Power Analysis "
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet:STATe
<State>
Activates a level offset at the sensor input. Define the appropriate value with the com-
mand
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SENS2:SWE:FREQ:OFFS:STAT ON
activates the specified level offset.
Options:
Option R&S SMF-K28
Manual operation:
See
" Level Offset State- Power Analysis "
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge[:STATe]
<State>
Activates the use of a frequency range for the power measurement that is different to
the set signal generator frequency range. The separate frequency range is entered
with commands
SENS:SWE:FREQ:SENS:STAR
and
SENS:SWE:FREQ:SENS:STop
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SENS2:SWE:FREQ:SENS:SRAN ON
activates use of a separate frequency range for frequency ver-
sus power measurement for sensor 2.
SENS2:SWE:FREQ:SENS:STAR 2.0GHZ
sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
SENS2:SWE:FREQ:SENS:STOP 2.9GHZ
sets a sweep stop at 2.9 GHz irrespective of the current signal
generator frequency settings.
Options:
Option R&S SMF-K28
Manual operation:
See
" Use Separate Frequency- Power Analysis "
Power Sensor Measurement Subsystems