Operation and handling
First value suppression
After a limit value has been exceeded, a reference measurement is carried out every 4 minutes
to evaluate the result. This prevents false alarms due to the counter ion effect after a system
shutdown.
The LIMESBase and the LIMESPlus are factory-equipped with a fixed activated first value
suppression.
Trigger conditions for an analysis start
•
Automatically by the set analysis interval
•
Manual analysis by pressing the "Start" button
•
3 minutes after switching on the unit
•
4 minutes after a limit value has been exceeded
•
Switching on a connected flow monitor after system standstill
•
Through a connected external control