Electrical installation
Connection input contact
Input contact (INPUT, start or stop analysis interval)
Terminal 13/12 - Black 2, Black 1
Potentially free flow monitors, switches or controls can be connected to the input contact to
start or pause analyses or to run an analysis interval. No external voltage may be applied to the
input contact.
For information regarding the wiring of the input contact, see page 33
Information regarding the function of the input contact depending on the programme switch can be
found from page 41
It is recommended to connect the input contact accordingly to avoid unnecessary
messages of a limit value exceeding during a regeneration.
Only connect potential-free switches to the terminals. Connecting an external voltage
source may damage the unit.