Electrical installation
Wiring example 4 and 5: Input contact INPUT (LIMESPlus only)
Potentially free flow monitors, switches or controls can be connected to the input contact to
start or pause analyses or to run an analysis interval. No external voltage may be applied to the
input contact.
Information regarding the function of the input contact depending on the programme switch can be
found from page 41
It is recommended to connect the input contact
accordingly to avoid unnecessary messages of a
limit value exceeding during a regeneration.
Only connect potential-free switches to the
terminals. Connecting an external voltage
source may damage the unit.
Voltage
LIMES
Plus
Jun
ct
ion
box
(o
pt
ion
al)
Illustration 17: Circuit diagram 4 and 5 - LIMESPlus