Operation and handling
Analysis interval LIMESPlus depending on the input contact INPUT
There are 4 fixed interval times that are set via the programme switches SW1 and SW2.
The interval time is the time between 2 consecutive analysis starts. If the input contact is open, no
analyses are started in the set analysis interval. Therefore, this input is bridged in the delivery state.
The bridge or an external switch (e.g., flow monitor) must be connected and closed when the water
is flowing in order to carry out analyses in the analysis interval.
Symbols used:
INPUT
Input contact
State
Description
Closed
The input contact is permanently closed
Open
The input contact is permanently open
Pulse
Only one analysis is carried out
Setting for the use of a switch or flow monitor
Internal analysis interval time,
input contact deactivated (SW3 OFF)
SW programme
switch position
INPUT
Input contact
Function
Analyses are carried out at
5 min intervals
Paused: No analyses are carried out because the
input contact is open.
Only one analysis is carried out
Analyses are carried out at
10 min intervals
Paused: No analyses are carried out because the
input contact is open.
Only one analysis is carried out
Analyses are carried out at
20 min intervals
Paused: No analyses are carried out because the
input contact is open.
Only one analysis is carried out
Analyses are carried out at
30 min intervals
Paused: No analyses are carried out because the
input contact is open.
Only one analysis is carried out