Operation and handling
Device settings
The LIMESPlus analyser is programmed via small slide switches (programme switches SW1 -
SW4) and adapted to the operating requirements. (SW = switch/switch)
Disconnect the unit from the power supply and remove the rubber cover of the
analyser.
The programme switches are located behind the rubber cover. The 16-step rotary switch for setting
the rinse time and the programme switches SW are located on this board.
Setting
Factory setting
Flushing time
Position 6
4 minutes
Setting
Factory setting
Analysis interval
SW1 OFF
SW2 ON
10 minutes
Input contact
SW3 OFF
From
Diagnostic mode
SW4 OFF
From
Work on electrical connections may only be carried out by authorised specialist
personnel in compliance with the locally applicable regulations.
You need a small screwdriver to operate the rotary switches and to set the
programme switches SW (slide switches). Please use only suitable and tested tools to
avoid damage to the sensitive components.
Voltage
Voltage